Update
4
4
X
Instrument
JammLite 4.3
JAMSXXXX.BIN
E
xit
A
XX.CLW
Port
I
P Address
Application
CLEW
I-STAT 300 Analyzer
COM1
Update
4
4
X
Instrument
JammLite 4.3
JAMSXXXX.BIN
Exit
AXX.CLW
Port
IP Address
Application
CLEW
I-STAT 300 Analyzer
COM1
If no ports are displayed, close
all open programs including
JammLite, and then re-launch
J
ammLite.
If JammLite still has no available
COM ports listed, call APOC
T
echnical Support for assistance.
By default, the lowest numbered COM port will
automatically be selected.
If the downloader is connected to a different COM
port, change the selection to that COM port now.
Art: 725708-00E Rev. Date 07-FEB-12 8
Check that the Application and CLEW listings
match those in the Product Update.
Note: Application and CLEW numbers are for example only.
The “numbers” have been replaced with X’s in the example
above and will change with each software update.
Click the Update button.
If an error occurs, check the
serial connection between the
downloader and the PC, as well
as the power connection to the
downloader.
If connected correctly, select a
different COM port (Do not select
TCP/IP) within the dropdown
menu and click Update.
If after trying each of the COM
ports listed in JammLite, you are
still not prompted to power off
the i-STAT 1 Handheld and insert
it into the downloader, call APOC
Technical Support for assistance.
1.13