Art: 725768-00E Rev. Date: 04-Jan-12
2
1. UPDATE PROCEDURE USING A NETWORK DOWNLOADER AND THE
JammLite PROCESS WITH TCP/IP
1.1: Before starting the process, make sure all the required equipment / informa-
tion is available:
• Computer with:
o Windows 2000, XP, or Windows 7 (32 bit)
o A CD-ROM drive
• i-STAT System Equipment
(1) Network Downloader, Downloader / Recharger, or DRC-300
Note: These instructions assume that the Network Downloader types used for
the update process are already installed and in-use on the customer’s network.
(2) i-STAT 1 handheld
(3) The latest i-STAT Software and Documentation CD
(4) Electronic Simulator
• List of the IP Address(es) for the network downloader(s) to be used for the soft-
ware update process
1.2: Close all open programs on the computer including i-STAT Central Data
Station (if applicable).
1.3: Before inserting the i-STAT 1 Software and Documentation CD, check that
the JAMS application and CLEW match the Product Update.
1.4: Insert the Software and Documentation CD into the CD-ROM drive.
• Wait 30 seconds, and then close any windows that may appear
• To access the files on the Software and Documentation CD-ROM:
o Double click the following icon
o Double click CD-ROM Drive

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals