9-4 Art: 714371-00E Rev. Date: 06-Aug-12
PREFERENCE WINDOW: for operator and patIent Id optIons
Option Description Default Comments
OPERATOR ID Minimum and maximum allowed operator ID length (scanned
or manually entered)
Min = 0 Max = 15 If operator IDs are a fixed length, the min. and max. settings
should both be equal to the ID length.
REPEAT ID ENTRY Operator must enter ID twice. Handheld prompts operator to
start again if IDs do not match.
Enabled: repeat required This option can be set for manual and/or scanned ID Entry.
INCLUDE ID ON
PRINTOUT
Enables/Disables printing of operator IDs on printouts from
the Martel or i-STAT printer.
Enabled Disabling the printing of operator IDs can prevent uncertified
operators from learning the IDs of certified operators.
BARCODE OPTIONS The type of barcodes used for Operator ID. See table below. All barcode types
MANUAL ENTRY CHECK
DIGIT
Options are None, ISBN Modulus 11 Check, and IBM
Modulus 10 Check.
None Check digit algorithms are given in HL7 Specification, Section
2.9.5.3
INVALID OPERATOR Behavior of handheld when Operator ID not in stored list or
certification date expired Options are: Not enabled (continue
without warning), Warn User (prompt to continue), and
Lockout (block testing until a valid Operator ID is scanned/
entered).
Continue without
warning
This option should not be enabled if the Use Operator List
option is disabled.
Separate Actions can be chosen for Certification Expired or
Operator Not On List.
EXPIRATION
NOTIFICATION
Allows a System Administrator to define a time period (1-255
days) in which the operator will be notified by a message on
the i-STAT 1 handheld display of their competency expiration
date.
Off See the Technoical Bulletin "Operator Competency Notification
on the i-STAT 1 Handheld" for full details.
PATIENT ID Minimum and maximum allowed patient ID length (scanned
or manually entered)
Min = 0 Max = 15 If ID numbers are a fixed length, the min. and max. settings
should both be equal to the ID length.
REPEAT ID ENTRY Operator must enter patient ID twice. Analyzer prompts
operator to start again if IDs do not match.
Repeat ID enabled This option can be set for manual and/or scanned ID entry.
PATIENT ID RECALL Operator can recall last patient ID when analyzer prompts for
Patient ID.
Enabled The most recent patient ID is recalled by pressing the key.
BARCODE OPTIONS The type of barcodes used for Patient ID. See table below. All barcode types
MANUAL ENTRY CHECK
DIGIT
Options are None, ISBN Modulus 11 Check, and IBM
Modulus 10 Check.
None Check digit algorithms are given in HL7 Specification, Section
2.9.5.3

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals