Alarms 79
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
32.
etCO
2
module
disconnected
Sensor alarms
CO
2
module is not
connected
Connect etCO
2
module
I
5
33.
etCO
2
FilterLine
disconnected
Sensor alarms
CO
2
filter line is not
connected
Connect filter line
I
5
34.
check etCO
2
sampling line
Sensor alarms
Malfunctioning CO
2
sampling line
Clean / replace
sampling line.
I
5
35.
check etCO
2
airway adapter
Sensor alarms
Malfunctioning CO
2
airway adapter
Clean / replace airway
adapter
I
5
36.
etCO
2
sensor
faulty
Sensor alarms
CO
2
sensor is faulty
Replace sensor or
Contact your local
ACUTRONIC Medical
Systems Distributor.
I
5
37.
SpO
2
module
disconnected
Sensor alarms
SpO
2
module is not
connected
Connect SpO
2
module
I
5
38.
SpO
2
sensor
failure (XY)
Sensor alarms
XY internal reference:
- 10: SpO
2
sensor is
faulty
- 38: SpO
2
module
DSP software
version not
supported
- 62: SpO
2
board
failure
Replace sensor or
disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor.
I
5
39.
check SpO
2
sensor
Sensor alarms
SPO
2
sensor is off
patient or
malfunctioning
Connect SpO
2
sensor
to patient
I
5
40.
Apnea alarm
Patient alarms
No spontaneous
patient breathing
detected.
Check patient,
Switch to controlled
ventilation.
I
6
41.
low SIQ
Patient alarms
SpO
2
sensor has low
SIQe (Signal Quality)
Check SpO
2
sensor for
correct contact with
patient or replace
sensor
I
7
42.
FiO
2
at max
Patient alarms
FiO
2
has reached
maximum setting
Adjust settings
I
7
43.
Lower
minimum FiO
2
Patient alarms
Lower minimum FIO
2
than set for PRICO
Adjust settings
I
7
44.
Low Pmean
Patient alarms
Leak or disconnection.
Check patient
connection.
Check tubing
system for leaks.
Check setting of
Pmean.
I
7

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals