80 Alarms
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
45.
High Pmean
Patient alarms
Pressure rise in tubing
system.
Check patient,
Check tubing
system,
Check exhalation
valve.
Replace patient
system.
I
7
46.
High PIP
Patient alarms
Pressure rise in tubing
system; mechanical
inspiration was
reduced to relieve the
system.
Check patient,
Check tubing
system.
Replace patient
system.
I
7
47.
High PEEP
Patient alarms
PEEP 6 mbar above
set value for at least 15
seconds.
Check patient,
Check patient
connection,
Check tubing.
Adjust settings
I
7
48.
High etCO
2
Patient alarms
End tidal CO
2
above
limit
Check patient,
Adjust settings
II
8
49.
Low etCO
2
Patient alarms
End tidal CO
2
below
limit
Check patient,
Adjust settings
II
8
50.
High FiCO
2
Patient alarms
Inspired CO
2
above
limit
Check patient,
Check patient dead
space,
Adjust settings
II
8
51.
Low FiCO
2
Patient alarms
Inspired CO
2
below
limit
Adjust settings
II
8
52.
High SpO
2
Patient alarms
SpO
2
measured
above set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
53.
Low SpO
2
Patient alarms
SpO
2
measured below
set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
54.
High Pulse
Rate
Patient alarms
Pulse rate measured
above set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
55.
Low Pulse
Rate
Patient alarms
Pulse rate measured
below set limit
Check patient,
II
8