4. Task start
Click to start the task process:
If connecting an external pressure module the correct standard should be selected
Click on the right to zero the pressure of the pressure module.
Click the task start button at the bottom right, and select the execution mode as automatic
execution or manual execution.
1) Automatic execution: The default measurement value is the same as the set point value. The
measured value can be manually changed.
2) Manual execution: Every time a set point is reached, the value will need to be entered before
advancing to the next set point.
The display icon to the right will change the display from graphical to numeric.
Standard process of calibration
1) When the calibration point is reached and stable, enter the measurement value for the analog
pressure gauge.
2) Click the Enter key on the keyboard or in the lower right corner of the screen to advance to the
next calibration point. At this time, the calibrator will start to control the pressure. The automatic
execution mode can be performed without manually advancing, if the measurement value is the
same as the set point value. ★ When enabling the tapping function, pay attention to the status
before/after tapping as prompted by the function screen, and input the indicated value of the
gauge before/after tapping.