128
automatically revert to the last setting.
Set point
1) There is no need to set this parameter when selecting the external pressure module.
2) It shall not exceed the selected measuring range.
3) Use this setting as the leak detection pressure point.
4) The pressure control set point value during the test.
Set the waiting time
1) Select the internal range: After the control pressure is stable during the execution (the real-time
pressure value turns green) and it switches to measure mode and starts the countdown of the waiting
time.
2) Select the external pressure module: Waiting time will be counted down after execution begins.
3) Enter the test time after the waiting time ends.
Set the test time
1) Select the internal range: After entering the test time, start the countdown, implement the test,
and stop the test after the time reaches 0.
2) Select the external pressure module: Same as the internal range.
Pressure type
1) Some devices can switch the gauge pressure and differential pressure.
2. Execution
Internal range