EasyManua.ls Logo

Agilent Technologies 1100 Series - Page 96

Agilent Technologies 1100 Series
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
96 1100 Series DAD and MWD User Manual
5 How to optimize the Detector
A wide bandwidth has the advantage of reducing noise by averaging over a
wavelength range compared to a 4 nm bandwidth, the baseline noise is
reduced by a factor of approximately 2.5, whereas the signal is about 75 % of a
4-nm wide band. The signal-to-noise ratio for a 30 nm bandwidth is twice that
for a 4-nm bandwidth in our example.
Because the detector averages absorbance values that are calculated for each
wavelength, using a wide bandwidth does not negatively impact linearity.
The use of a reference wavelength is highly recommended to further reduce
baseline drift and wander induced by room temperature fluctuations or
refractive index changes during a gradient.
An example of the reduction of baseline drifts is shown in Figure 56 for
PTH-amino acids. Without a reference wavelength, the chromatogram drifts
downwards due to refractive index changes induced by the gradient. This is
almost completely eliminated by using a reference wavelength. With this
technique, PTH-amino acids can be quantified in the low picomole range even
in a gradient analysis.
Figure 55 Influence of Bandwidth on Signal and Noise
Bandwidth
30 nm
12 nm
4 nm

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals