EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>1100 Series

Agilent Technologies 1100 Series User Manual

Agilent Technologies 1100 Series
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
1100 Series DAD and MWD User Manual 27
Installing the Detector 2
Flow Connections to the Detector
Preparations Detector is installed in the LC system.
Parts required Other modules
Parts from accessory kit, see“Detector Accessory Kit Contents” on page 21.
Two wrenches 1/4 5/16 inch for capillary connections
WARNING
When working with solvents please observe appropriate safety procedures (for
example, goggles, safety gloves and protective clothing) as described in the
material handling and safety data sheet supplied by the solvent vendor, especially
when toxic or hazardous solvents are used.
NOTE
The flow cell is shipped with a filling of isopropanol (also recommended when the
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
NOTE
The detector should be operated with the front cover in place to protect the flow cell area
against strong drafts from the outside and to cover the deuterium lamp.
Some types of the Agilent deuterium lamps show a light ring during operation. This is not
harmful, refer to “UV-Radiation” on page 204.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals