1100 Series DAD and MWD User Manual 27
Installing the Detector 2
Flow Connections to the Detector
Preparations Detector is installed in the LC system.
Parts required Other modules
Parts from accessory kit, see“Detector Accessory Kit Contents” on page 21.
Two wrenches 1/4 – 5/16 inch for capillary connections
WARNING
When working with solvents please observe appropriate safety procedures (for 
example, goggles, safety gloves and protective clothing) as described in the 
material handling and safety data sheet supplied by the solvent vendor, especially 
when toxic or hazardous solvents are used.
NOTE
The flow cell is shipped with a filling of isopropanol (also recommended when the 
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to 
subambient conditions.
NOTE
The detector should be operated with the front cover in place to protect the flow cell area 
against strong drafts from the outside and to cover the deuterium lamp.
Some types of the Agilent deuterium lamps show a light ring during operation. This is not 
harmful, refer to “UV-Radiation” on page 204.