100 1100 Series DAD and MWD User Manual
5 How to optimize the Detector
Optimizing Spectral Acquisition (DAD only)
Storage of all spectra consumes a lot of disk space. It is very useful to have all
spectra available during optimization of a method or when analyzing unique
samples. However when running many samples of the same type, the large size
of data files with all spectra may become a burden. The detector provides
functions to reduce the amount of data, yet retaining the relevant spectral
information.
For spectra options see “Spectrum Settings” on page 81.
Range
Only the wavelength range where the compounds in your sample absorb
contains information that is useful for purity checks and library searches.
Reducing the spectrum storage range saves disk space.
Step
Most substances have broad absorbance bands. Display of spectra, peak purity
and library search works best if a spectrum contains 5 to 10 data points per
width of the absorbance bands. For anisic acid (the example used before) a
step of 4 nm would be sufficient. However a step of 2 nm gives a more optimal
display of the spectrum.
Threshold
Sets the peak detector. Only spectra from peaks higher than threshold will be
stored when a peak-controlled storage mode is selected.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals