124 1100 Series DAD and MWD User Manual
8Maintenance
Repairing the Standard or Semi-Micro Flow Cell
When required If the flow cell needs repair due to leaks or contaminations (reduced light
throughput)
Tools required Two 1/4 inch wrenches for capillary connections
hexagonal key 4 mm
Tooth picks
Parts required For parts, see “Standard Flow Cell” on page 156 and “Semi-Micro Flow
Cell” on page 158
Preparations Turn the flow off.
Remove the front cover.
Remove the flow cell, see “Exchanging a Flow Cell” on page 121.
NOTE
The gaskets used in the standard and semi-micro flow cell are different.