1100 Series DAD and MWD User Manual 45
LAN Configuration 3
Storing the settings permanently with Bootp
If you want to change parameters of the detector using the Bootp follow the
instructions below.
1 Turn off the detector.
2 Change the detector’s settings of the Configuration Switch to “Bootp &
Store” mode, see Table 6 on page 35.
3 Start the Agilent Bootp Service and open its window.
4 If required, modify the parameters for the detector according to your needs
using the existing configuration.
5 Press OK to exit the Bootp Manager.
6 Now turn on the detector and view the Bootp Server window. After some
time the Agilent Bootp Service will display the request from the LAN
interface. The parameters are now stored permanently in the non-volatile
memory of the detector.
7 Close the Agilent Bootp Service and turn off the detector.
8 Change the settings of the detector’s Configuration Switch to “Using
Stored” mode, see Table 6 on page 35.
9 Power cycle the detector. The detector can be accessed now via LAN
without the Agilent Bootp Service, refer to “PC and Agilent ChemStation
Setup” on page 54.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals