1100 Series DAD and MWD User Manual 35
LAN Configuration 3
Initialization mode selection
The following initialization (init) modes are selectable:
Bootp
When the initialization mode “Bootp” is selected, the detector tries to
download the parameters from a Bootp Server. The parameters obtained
become the active parameters immediately. They are not stored to the
non-volatile memory of the detector. Therefore, the parameters are lost with
the next power cycle of the detector.
Figure 8 Bootp (Principle)
Bootp & Store
When “Bootp & Store” is selected, the parameters obtained from a Bootp
Server become the active parameters immediately. In addition, they are stored
to the non-volatile memory of the detector. Thus, after a power cycle they are
still available. This enables a kind of “bootp once” configuration of the
detector.
Table 6 Initialization Mode Switches
SW 6SW 7SW 8Init Mode
OFF OFF OFF Bootp
OFF OFF ON Bootp & Store
OFF ON OFF Using Stored
OFF ON ON Using Default
1
2
345678
ON
Bootp
Server
Active
Parameter