1100 Series DAD and MWD User Manual 123
Maintenance 8
6
Insert the flow cell while pressing the flow cell holder. 7 Insert the flow cell capillaries into the union holder (top is
inlet, bottom is outlet). Tighten the thumb screw and
Reconnect the waste tubing (bottom) to the union.
Note:
To check for leaks, establish a flow and observe the
flow cell (outside of the cell compartment) and all
capillary connections.
Next steps:
8
Perform a “Wavelength Verification and Recalibration” as
described in the Service Manual to check the correct
positioning of the flow cell.
9 Replace the front cover.
inlet
outlet
thumbscrew
waste