Agilent 1260 Infinity DAD and MWD User Manual 109
How to optimize the Detector
5
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Because the detector averages absorbance values that are calculated for each
wavelength, using a wide bandwidth does not negatively impact linearity.
The use of a reference wavelength is highly recommended to further reduce
baseline drift and wander induced by room temperature fluctuations or
refractive index changes during a gradient.
An example of the reduction of baseline drifts is shown in Figure 37 on
page 109 for PTH-amino acids. Without a reference wavelength, the
chromatogram drifts downwards due to refractive index changes induced by
the gradient. This is almost completely eliminated by using a reference
wavelength. With this technique, PTH-amino acids can be quantified in the low
picomole range even in a gradient analysis.
Figure 37 Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without
Reference
7LPHPLQ
*UDGP.+32$&1IURP$&1WR$&1LQPLQ
SPROHDFK
5HIHUHQFHQP
:DYHOHQJWK
QP
:DYHOHQJWK
QP
1RUHIHUHQFH
37+$61
37+$5*
37+$/$
37+3+(
37+352


Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315D and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315D Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315D
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals