58 Agilent 1260 Infinity DAD and MWD User Manual
3
Installing the Module
Installing Capillaries
5 When using UHP-FF fittings with bioinert capillaries, do not try to remove
fittings from these capillaries. Bio-inert capillaries are using a PEEK front
end, which may expand under pressure especially when being in contact
with some organic solvents. If a fitting is moved across an expanded PEEK
end, there is a risk of damaging the capillary by ripping off its end. Before
re-installing such capillaries, push the ferrule towards the rear site for a
small distance.
Figure 13 Capillary fitting