General Troubleshooting 2
5977B Series MSD Troubleshooting and Maintenance Manual 41
Very dirty ion source
Calibration valve is not working correctly
Bad signal cable connection
Filament has failed, or is not connected correctly
Bad ion source wiring connection
Bad detector wiring connection
Failed electron multiplier horn
Peaks are tailing
Active sites in the sample path
Injection is too large
Incorrect GC inlet temperature
Insufficient column flow
GC/MSD interface temperature is too low
Ion source temperature is too low
Peaks are fronting
Column film thickness mismatched with analyte concentration
(column overload)
Initial oven temperature is too low
Active sites in the sample path
Injection is too large
GC inlet pressure too high
Insufficient column flow
Peaks have flat tops
Insufficient solvent delay
Incorrect scale on the display
Injection is too large
Electron multiplier voltage is too high
Gain is too high

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals