102 Maintaining Your GC
6 Maintaining the Packed Column Inlet
To Change the Glass Liner on the Packed Column Inlet
1 Gather the following:
Replacement glass liner (See “Consumables and Parts for
the Packed Column Inlet” on page 92.)
9/16-inch wrench
Lint-free gloves
2 Load the inlet maintenance method and wait for the GC to
become ready.
3 Loosen the top insert weldment to remove the top portion of
the inlet.
4 Use a thin wire or wood splint to carefully lift and remove
the old glass liner.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Top insert weldment
Flared end
Glass liner
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.