214 Maintaining Your GC
11 Maintaining the NPD
To Attach a Capillary Column Adapter on an Adaptable NPD
1 Gather the following materials:
Adapter (See “Consumables and Parts for the NPD” on
page 208.)
1/4-inch nut
1/4-inch ferrule
Column cutter
1/4-inch wrench
9/16-inch open-end wrench
Lint-free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Assemble a brass nut and Vespel/graphite ferrule onto the
adapter.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
1/4-inch ferrule
1/4-inch nut
Adapter

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals