Maintaining the COC Inlet 7
Maintaining Your GC 143
To Bakeout Contaminants from the COC Inlet
1 Set the column flow to the normal operating value, or set the
capillary column gas velocity to 30 cm/s.
2 Purge the column with carrier flow for at least 10 minutes
before heating the oven.
3 Set the inlet mode to Oven Track.
4 If the column is attached to the detector, set the detector
25 °C above normal operating temperature.
If the column is not attached to the detector, cap the detector
fitting.
5 Set the column oven 25 °C above the GC method final oven
temperature to bake contaminants out of the inlet. Do not
exceed the column manufacturer’s maximum temperature
limit.
6 Bakeout for 30 minutes or until the detector baseline is free
of contamination peaks.