Maintaining the Purged Packed Inlet 5
Maintaining Your GC 77
To Install an Insulation Cup on the Purged Packed Inlet
1 Gather the following:
No-hole ferrule
Column nut
2 Install a plug (for example, a column nut with no-hole
ferrule) in the inlet capillary adapter.
3 Push the cup spring to the right. Slide the cup over the inlet
fitting so that the insulation at the top of the cup is flush
against the oven roof.
4 Place the spring into the groove in the inlet liner. Remove the
column nut and no-hole ferrule.