288 Maintaining Your GC
A Swagelok Connections
Making Swagelok Connections
Objective
To make a tubing connection that does not leak and that can be
taken apart without damaging the fitting
Materials needed:
1/8-inch (or 1/4-inch, if used) preconditioned copper tubing
1/8-inch (or 1/4-inch, if used) Swagelok nuts
Front and back ferrules
Two 7/16-inch (for 1/8-inch nuts) or 9/16-inch (for 1/4-inch
nuts) wrenches
1 Place a Swagelok nut, back ferrule, and front ferrule to the
tubing as shown in Figure 8.
Figure 8 Swagelok nuts and ferrules
2 Clamp a stainless steel plug or similar fitting in a bench vise.
3 Push the tubing into the stainless steel plug (Figure 9).
4 Make sure that the front ferrule is touching the plug. Slide
the Swagelok nut over the ferrule and thread it onto the plug.
Front ferrule
Back ferrule
Nut
Tubing
IMPORTANT!
The narrow end of the back
ferrule fits into the rear of
the front ferrule.
CAUTION
Do not
Use a separate stainless steel fitting in a vise for initial tightening of
the nut. Do not use an inlet or detector fitting. Strong forces are
required to properly set the ferrules, and damage to an inlet or
detector fitting is very costly to repair.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals