EasyManuals Logo

Agilent Technologies 7820A User Manual

Agilent Technologies 7820A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #200 background imageLoading...
Page #200 background image
200 Maintaining Your GC
10 Maintaining the µECD
To Install a Capillary Column in the µECD
1 Gather the following:
Ferrule (See “Consumables and Parts for the µECD” on
page 194.)
Column nut
Septum
Column
1/4-inch, 5/16-inch, and 9/16-inch wrenches
Column cutter
Lint-free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Load the inlet maintenance method and wait for the GC to
become ready.
4 Place a septum, capillary column nut, and ferrule on the
column.
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals