48 Maintaining Your GC
4 Maintaining the Split/Splitless Inlet
To Change the Liner and O-Ring on the Split/Splitless Inlet
1 Gather the following:
Replacement O-ring (See “Consumables and Parts for the
Split/Splitless Inlet” on page 36.)
Replacement liner
Tweezers
Lint-free gloves
2 Load the inlet maintenance method and wait for the GC to
become ready.
3 Unscrew the insert nut from the inlet body. Lift the septum
assembly straight up and away from the inlet to avoid
chipping or breaking the liner.
If using a Flip-Top inlet, lift the lever arm.
4 Loosen the O-ring from the sealing surface with tweezers.
5 Grasp the liner with tweezers and pull it out.
6 Inspect the surface of the gold seal for graphite or rubber
septum contamination. If required, replace the gold seal.
(See “To Replace the Gold Seal on the Split/Splitless Inlet” on
page 50.)
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Liner and O-ring in inlet

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals