EasyManua.ls Logo

Agilent Technologies 7820A - To Install a Capillary Column in the TCD

Agilent Technologies 7820A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
184 Maintaining Your GC
9 Maintaining the TCD
To Install a Capillary Column in the TCD
1 Gather the following:
Front ferrule (See “Consumables and Parts for the TCD”
on page 182.)
Back ferrule
Column nut
Column cutter
7/16-inch wrench
Lab tissue
Lint-free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Assemble the ferrules and 1/8-inch brass Swagelok nut on
the column.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Front ferrule
Back ferrule
Nut

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals