248 Maintaining Your GC
12 Maintaining the FPD
+
To Install the FPD
+
Cover
1 Gather the following:
T-20 Torx screwdriver
2 Install the cover.
Single-wavelength detector:
a Start the two screws on the right side of the cover.
b Start and tighten the screws at the base on the left side.
c Tighten the screws on the right side.
Dual-wavelength detector: Install the cover (two screws).
3 Close the FPD
+
detector top cover.