Maintaining the FID 8
Maintaining Your GC 151
Selecting an FID Jet
Open the oven door and locate the column connection fitting at
the base of the detector. It will look like either a capillary
optimized fitting or an adaptable fitting.
• If you have an application that tends to clog the jet, select a
jet with a wider tip id.
• When using packed columns in high column-bleed
applications, the jet tends to clog with silicon dioxide.
• In simulated distillation applications, the high-boiling
hydrocarbons tend to clog the jet.
For capillary optimized fittings, select a jet from Table 24. For
adaptable fittings, select a jet from Table 25.
Capillary optimized fitting
Adaptable fitting
Detector fitting
Adapter
Table 24 Jets for capillary optimized fittings
Figure 1 ID Jet type Part number Jet tip id Length
1 Capillary G1531-80560 0.29 mm (0.011 inch) 43 mm
2 High-temperature
(use with simulated
distillation)
G1531-80620 0.47 mm (0.018 inch) 48 mm
Figure 2 Capillary optimized jets