244 Maintaining Your GC
12 Maintaining the FPD
+
To Remove the FPD
+
Cover
1 Gather the following:
T-20 Torx screwdriver
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Open the FPD
+
detector top cover.
4 Loosen the screws securing the FPD
+
cover to the top of the
detector.
CAUTION
When turning off the GC, first turn off the flame to prevent
condensation from dripping into the jet and column.
WARNING
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.

Table of Contents