28 Maintaining Your GC
3 Maintaining Capillary Columns
To Condition a Capillary Column
1 Gather the following:
One 7/16-inch, and 1/4-inch wrenches
No-hole ferrule (See “Consumables and Parts for
Columns” on page 24.)
Column nut
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Install the column into the inlet using the new ferrules. See:
To Install a Capillary Column with the Split/Splitless Inlet
To Install a Capillary Column with the Purged Packed
Inlet
To Install a Capillary Column with the COC Inlet
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals