130 Maintaining Your GC
7 Maintaining the COC Inlet
To Change a Septum on the COC Inlet
1 Gather the following:
Replacement septum. (See “Consumables and Parts for
the COC Inlet” on page 120.)
Tweezers
A thin wire (0.2-inch diameter) for removing septum from
inlet
Lint-free gloves
2 Load the inlet maintenance method and wait for the GC to
become ready.
3 Replace the septum.
If you are using a septum nut, grasp the knurling and
unscrew. Remove the old septum with tweezers. Use
tweezers to install a new septum. Push the septum into
the septum nut until properly seated. Firmly tighten the
nut.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
For 250/320-µm
automated injections
For 530-µm automated
injections

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals