158 Maintaining Your GC
8 Maintaining the FID
To Replace the FID Collector Assembly
1 Gather the following:
New FID collector assembly. (See “Consumables and Parts
for the FID” on page 146.)
T-20 Torx screwdriver
1/4-inch nut driver
Tweezers
Lint-free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Disconnect the ignitor cable assembly.
4 Remove the three screws holding the collector assembly to
the mounting pallet.
CAUTION
To avoid contaminating the FID, wear clean, lint-free gloves when
handling the collector assembly.