Maintaining the Split/Splitless Inlet 4
Maintaining Your GC 55
To Clean the Split/Splitless Inlet
1 Gather the following:
Replacement septum (See “Consumables and Parts for the
Split/Splitless Inlet” on page 36.)
Replacement liner
Replacement O-ring
Replacement gold seal
Replacement washer
Solvent that will clean the type of deposits in your inlet
Compressed, filtered, dry air or nitrogen
Beaker
Cleaning brushes—The FID cleaning kit (part number
9301-0985) contains appropriate brushes
Lint-free gloves
2 Load the inlet maintenance method and wait for the GC to
become ready.
3 Remove the inlet liner. (See “To Change the Liner and O-Ring
on the Split/Splitless Inlet” on page 48.)
4 Disconnect the column from the inlet.
5 Remove the reducing nut and gold seal. (See “To Replace the
Gold Seal on the Split/Splitless Inlet” on page 50.)
6 Place a beaker in the oven under the inlet to catch the
solvent.
7 Soak a cleaning brush in the solvent and scrub the inside of
the inlet weldment. Repeat 10 times.
8 Rinse the inlet with the solvent.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals