About Maintaining the GC 1
Maintaining Your GC 13
Maintenance Methods for the 7820A GC
Before most maintenance procedures, the GC must be made 
ready. Agilent recommends that you create and store the 
following maintenance methods into the GC. The methods below 
will:
• Prevent damage to the instrument (electronics, columns, 
etc.)
• Avoid injury to the user (burns, shocks, etc.)
• Allow you to perform maintenance on specific areas while 
leaving the rest of the GC components at operating 
temperature
You can use the software keyboard or your Agilent data system 
to create, save, and load these methods.
General GC Maintenance Method
Create this method for instrument column maintenance, 
detector maintenance, and general GC maintenance tasks. 
• Set the oven temperature to 35 °C. This allows the oven fan to 
assist cooling.
• Set all inlet temperatures to 35 °C and set inlet gas pressures 
to 0.0.
• If performing column maintenance, remember to wait for 
the oven and column to cool down before turning off 
column carrier gas flow at the source. Also remember to 
cap both ends of the column to keep air out once it is 
removed.
• If you are not performing column maintenance, keep inert 
carrier gas (helium or nitrogen) flowing to protect the 
column.
• Set all detector temperatures to 35 °C. 
• If performing FPD
+
 maintenance, turn off the GC and 
unplug the power cord.
• Some detectors (FID, NPD) use high voltages. For these 
detectors, turn the electrometer Off to disable the high 
voltage.
NOTE
Inlets and detectors at operating temperature may require 12 hours or 
longer to reach the maintenance method setpoints below.