Maintaining the G3435A/G3436A FPD 13
Maintaining Your GC 255
To Install a Capillary Column Adapter in the FPD
1 Gather the following:
FPD capillary column adapter (See “Consumables and
Parts for the FPD” on page 252.)
Column cutter
1/4-inch wrench
9/16-inch wrench
Metric ruler
1/8-inch nut
Lint-free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Insert the capillary adapter into the 1/8-inch nut as shown,
then thread the nut onto the detector fitting.
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals