EasyManua.ls Logo

Agilent Technologies 7820A - To Change the Septum on the Split;Splitless Inlet

Agilent Technologies 7820A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44 Maintaining Your GC
4 Maintaining the Split/Splitless Inlet
To Change the Septum on the Split/Splitless Inlet
1 Gather the following:
Replacement septum. (See “Consumables and Parts for
the Split/Splitless Inlet” on page 36.)
0- or 00-grade steel wool (optional)
Tweezers
2 Load the inlet maintenance method and wait for the GC to
become ready.
3 Remove the septum retainer nut or Merlin cap.
4 Use tweezers to remove the septum or Merlin Microseal from
the retainer nut. Do not gouge or scratch the interior of the
septum head.
5 Firmly press the new septum or Merlin Microseal into the
fitting. The metal parts side of the Merlin Microseal should
face down (toward the oven).
6 Install the septum retainer nut or Merlin cap and
finger-tighten. Tighten the septum retainer nut until the
C-ring is about 1 mm above the nut.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
SeptumSeptum retainer
nut

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals