Maintaining the µECD 10
Maintaining Your GC 203
To Install the Insulating Cup for the µECD
1 Gather the following:
Nut warmer insulation (See “Consumables and Parts for
the µECD” on page 194.)
Insulation cup assembly
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Install the insulation in the cup. Line up the slots in the
insulation with the slot in the cup.
4 Push the wire spring lever to the right to uncover the hole.
5 With a capillary column installed to prevent insulation
contamination of the detector, slide the column into the slot
on the insulation cup and place the cup over the makeup gas
adapter.
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.