EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7820A

Agilent Technologies 7820A User Manual

Agilent Technologies 7820A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #257 background imageLoading...
Page #257 background image
Maintaining the G3435A/G3436A FPD 13
Maintaining Your GC 257
To Attach a Capillary Column to the FPD
1 Gather the following:
Column measuring tool (See “Consumables and Parts for
the FPD” on page 252.)
Column cutter
1/4-inch and 7/16-inch wrenches
Column nut
Ferrule
Capillary column
Lint-free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Assemble a septum, column nut, and ferrule on the end of
the column.
4 Insert the end of the column through the column measuring
tool so that the end protrudes beyond the tool.
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals