40 Maintaining Your GC
4 Maintaining the Split/Splitless Inlet
To Install a Capillary Column with the Split/Splitless Inlet
1 Gather the following (see “Consumables and Parts for the
Split/Splitless Inlet” on page 36):
Column
Ferrule(s)
Column nut
Septum
Column cutter
Isopropanol
Lab tissue
Metric ruler
1/4-inch open-end wrench
Lint-free gloves
2 Load the GC maintenance method and wait for the GC to
become ready.
3 Verify that the correct glass liner is installed. (See
“Consumables and Parts for the Split/Splitless Inlet” on
page 36.)
4 Place the column on the hanger with the ends pointing up
and the label to the front.
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals