EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7820A

Agilent Technologies 7820A User Manual

Agilent Technologies 7820A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
74 Maintaining Your GC
5 Maintaining the Purged Packed Inlet
To Change the O-Ring on the Purged Packed Inlet
1 Gather the following:
Replacement O-ring (See “Consumables and Parts for the
Purged Packed Inlet” on page 60.)
Septum nut wrench
Tweezers
Lint-free gloves
2 Load the inlet maintenance method and wait for the GC to
become ready.
3 Loosen the top insert weldment to remove the top portion of
the inlet.
4 Use tweezers to remove the old O-ring.
5 Insert a new O-ring.
6 Install and tighten the top insert weldment.
7 Restore the analytical method.
8 Reset the EMF counter. See To Reset an EMF Counter in the
Operation Manual.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals