EasyManua.ls Logo

Agilent Technologies 7820A - Page 201

Agilent Technologies 7820A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintaining the µECD 10
Maintaining Your GC 201
5 Score the column using a glass scribing tool. The score must
be square to ensure a clean break.
6 Break off the column end by supporting it against the column
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
7 Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
8 Install the column.
If the column id is 200 um or more, push the column into
the adapter until it stops at the indentation. Pull it back
1 to 2 mm and tighten the column nut with one 5/16-inch
wrench on the adapter and another 1/4-inch wrench on the
column nut.
Ferrule
Column nut
Septum
Good
Bad

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals