76 Maintaining Your GC
5 Maintaining the Purged Packed Inlet
5
Wearing gloves, inspect the O-ring and replace, if necessary.
(See “To Change the O-Ring on the Purged Packed Inlet” on
page 74.)
6 Wearing gloves, grasp the flared end (top) of the replacement
glass liner with tweezers and install it in the inlet. If the glass
liner does not seat properly because a capillary column is
installed, remove the column, install the glass liner, and
replace the column. (See “To Install a Capillary Column with
the Purged Packed Inlet” on page 64.)
7 Install the top insert weldment and hand-tighten firmly.
8 Restore the analytical method.
9 Reset the EMF counter. See To Reset an EMF Counter in the
Operation Manual.