106 Maintaining Your GC
6 Maintaining the Packed Column Inlet
12
Install the adapter, if used. (See “To Install an Adapter on the
Packed Column Inlet” on page 99.)
13 Install the glass liner and O-ring. (See “To Change the Glass
Liner on the Packed Column Inlet” on page 102.)
14 Install the top insert weldment and finger-tighten.
15 Install the septum and septum nut. (See “To Change the
Septum on the Packed Column Inlet” on page 95.)
16 Attach the column.
17 Check for leaks.
18 Restore the analytical method.
19 Set the septum and glass liner counters.