EasyManua.ls Logo

Agilent Technologies 7820A - Page 126

Agilent Technologies 7820A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
126 Maintaining Your GC
7 Maintaining the COC Inlet
6
Break off the column end by supporting it against the column
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
7 Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
8 Gently insert the column into the inlet until it bottoms. You
should feel spring tension as you push up on the column. (Do
not withdraw the column.)
9 Insert the column nut into the inlet fitting and finger-tighten.
Good
Bad

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals