EasyManua.ls Logo

Agilent Technologies 7820A - Page 156

Agilent Technologies 7820A
294 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
156 Maintaining Your GC
8 Maintaining the FID
4
Place a septum, capillary column nut, and ferrule on the
column.
5 Score the column using a glass scribing tool. The score must
be square to ensure a clean break.
6 Break off the column end by supporting it against the column
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
7 Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
8 Install the capillary column.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Ferrule
Column nut
Septum
Good
Bad

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals