Maintaining the FID 8
Maintaining Your GC 163
9 Carefully screw the jet into the housing. Tighten 1/6-turn 
past finger-tight (1/6-turn is one “flat” on a typical 
screwdriver handle, or the jet head).
10 Install the collector assembly. (See “To Replace the FID 
Collector Assembly” on page 158.) 
11 Reset the jet counter. See To Reset an EMF Counter in the 
Operation Manual.
12 Attach the capillary column to the detector. 
a Install the column in the detector. (See “To Install a 
Capillary Column in the FID” on page 155.)
b After the column is installed at both inlet and detector, 
establish a flow of carrier gas and purge as recommended 
by the column manufacturer. 
c Check the FID leakage current. (See “To Check the FID 
Leakage Current” on page 173.)
d Bakeout the detector. (See “To Bakeout the FID” on 
page 178.)
e Restore the analytical method.
CAUTION
Do not overtighten the jet! Overtightening may permanently deform 
and damage the jet, the detector base, or both. The torque 
specification is 10 inch-pounds. 
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to 
cause burns. If the oven, inlet, or detector is hot, wear 
heat-resistant gloves to protect your hands.