166 Maintaining Your GC
8 Maintaining the FID
3
Remove the FID ignitor.
a Disconnect the ignitor cable assembly.
b Loosen the ignitor with a wrench.
c Turn the nut counterclockwise by hand. Remove the
ignitor and copper washer.
4 If replacing only the FID ignitor assembly with copper
washer, skip to step 16 for assembly.
5 Remove the three screws that hold the collector mount to the
FID thermal strap.