172 Maintaining Your GC
8 Maintaining the FID
17
Lower the collector assembly into the housing.
18 Insert the three screws and tighten (to 18 inch-pounds).
19 Connect the ignitor extension cable.
20 Verify assembly:
a Check the FID leakage current. (See “To Check the FID
Leakage Current” on page 173.)
b Bakeout the detector. (See “To Bakeout the FID” on
page 178.)
c Check the FID baseline. (See “To Check the FID Baseline”
on page 174.)
21 Reset the EMF counters. See To Reset an EMF Counter in the
Operation Manual.