Maintaining the G3435A/G3436A FPD 13
Maintaining Your GC 265
6 Remove the O-ring with tweezers.
7 Assemble the parts for the new ignitor.
8 Install the new ignitor assembly and tighten with a wrench.
Do not overtighten.
9 Replace the ignitor collar and cable assembly and tighten the
screw.
10 Replace the cover and the vent tube assembly. (See “To
Install the FPD Vent Tube and Cover” on page 266.)
11 Restore the analytical method.
12 Wait 20 minutes for the detector to heat up, then ignite the
flame.
13 Reset the EMF counter. See To Reset an EMF Counter in the
Operation Manual.
Glow plug
Spacer
O-ring