42 Maintaining Your GC
4 Maintaining the Split/Splitless Inlet
10
Thread the column nut into the inlet but do not tighten.
11 Adjust the column position so that the septum contacts the
bottom of the column nut. Finger-tighten the column nut
until it begins to grip the column.
12 Tighten the column nut an additional 1/4 to 1/2 turn with a
wrench so that the column cannot be pulled from the fitting
with gentle pressure.
13 Configure the new column.
14 Condition the column per the manufacturer’s
recommendation. (See To Condition a Capillary Column.)
15 Install the column into the detector. See:
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the µECD
To Attach a Capillary Column to the FPD
+
16 After the column is installed at both inlet and detector,
establish a flow of carrier gas and purge as recommended by
the column manufacturer.
17 Restore the analytical method.
For FID or FPD
+
, immediately turn off the flame.
For NPD, immediately turn off the bead.
4%6 mm

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals