EasyManua.ls Logo

Agilent Technologies 7820A - Page 65

Agilent Technologies 7820A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintaining the Purged Packed Inlet 5
Maintaining Your GC 65
6 Score the column using a glass scribing tool. The score must
be square to ensure a clean break.
7 Break off the column end by supporting it against the column
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
8 Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
9 Position the column so it extends 1 to 2 mm above the end of
the ferrule. Slide the septum up the column to hold the
column nut at this fixed position.
Ferrule
Column nut
Septum
Good
Bad

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals